VEGA II Installation Requirements:

  • Power 230 V/50 Hz, optionally 120 V/60 Hz, 1300 VA
  • No water cooling
  • Compressed dry nitrogen is recommended: 150–500 kPa
  • Compressed air for suspension: 450–600 kPa(only LM and XM chambers)
  • Room for installation: min. 3 m × 3 m

System dimension: 2.15 m × 1.075 m

Environmental requirements:

  • Temperature of environment 15–28°C
  • Relative humidity: <80 %
  • Background magnetic field:
    • synchronous: < 3 · 10-7 T;
    • asynchronous: < 1 · 10-7 T
  • Vibrations:
    • Passsive isolation (mechanic): <4 μm/s below 30 Hz; <8 μm/s above 30 Hz
    • Passive isolation (pneumatic): <6 μm/s below 30 Hz; <12 μm/s above 30 Hz
    • Active isolation – option: <12 μm/s below 30 Hz; <24 μm/s above 30 Hz